3.4.1 ALARM SETUP
Alarm Setup has the following options:
ALM REC TIME (Alarm Record Time):
In case of physiological alarm, the system can record the information before and after the alarm
time. The system can give three kinds of time, i.e., 4 seconds, 8 seconds, 16 seconds. The seconds
in the options are the sum of seconds before and after the alarm time. For example, 8 seconds
mean the information within 4 seconds before the alarm time and within 4 seconds after such a
time.
ALM PAUSE TIME (Alarm pause time)“1 minute”, “2 minutes” or “3 minutes”
ALM LIMIT (Display alarm limits)”ON” and”OFF”
ALM LATER: Disabled ,5s,10s,15s,20s. Select "Disable".
when an alarm occurs, the instrument that occurs immediately alarm; When Select "five seconds",
alarm occurs only when the parameters continue to exceed the the alarm upper or lower limit of 5
seconds instrument, other settings and so on. The purpose of this feature is to reduce the instant
false alarms, allow doctors to more accurately grasp the situation.
3.4.2 RECORD (Record output setting)
Options for record output:
REC WAVE1/ REC WAVE2: can not change
REC RATE: 25.0/50.0 mm/s, record speed
REC GRID: ON/OFF
RT REC TIME: 3S/5S/8S/CONTINUAL (real-time record time)
CONTINUAL means after pressing PRINT BUTTON
, monitor will keep printing
parameters and waveform, until press
to stop
Caution
The recorder is a component for option.
Caution
Where two similar waveforms are selected, the system will automatically adjust the
other one into a different one.