Chapter15 TEMP Monitoring
15.1 TEMP Monitoring
The monitor has two TEMP measuring channels. The temperature data can be measured with the
TEMP detector. Teperature is not a standard configuration for the neonatal monitor.
15.1.1 TEMP measurement setting
For one-time TEMP detectors, users must insert the TEMP cables into slots and then connect the
detectors with such cables; for reusable TEMP detectors, users can directly insert them into slots.
Closely paste TEMP detectors with patient body.
Turn on the system power supply.
Warning
Before monitoring users should check status of detector cables by plugging out the
TEMP detector cable from the hole, then the screen will display the error information
“T sensor disconnected” and make voice alarming.
Caution
A one-time TEM detector can only be used once.
Warning
Be careful to use or store TEMP detector and cables; spare detectors and cables should
be wrapped into loose rolls. Tough wires inside the detector and cables, if any, may
cause mechanical injury.
Warning
Calibration of a TEMP detector must be done for every two years or comply with your
hospital’s specified schedule. When requiring calibration, please contact the
manufacturer.
Caution
During monitoring process a TEMP detector will self-detect once per hour; such
self-detection last for 2 seconds and will not affect normal work of the TESP monitor.