3.4.3 TIME SETUP (System Time Setting)
"TIME SETUP" with options below:
Setup year, month, day, hours, minutes and seconds.
Pitch on the content required for correction.
3.4.4 3.4.4 NURSE CALL SETUP
Options in nurse call setting:
ALM ON/OFF: ON/OFF
ALM LEV(Alarm lever): high, middle, low
3.4.5 MODULE SETUP
Options in module setup:
ECG and SpO2 are not optional.
TEMP: ON/OFF. When it is ON, machine can monitor TEMP, when it is OFF, machine can
not monitor TEMP, and no parameter area of TEMP in main interface.
RESP: ON/OFF. When it is ON, machine can monitor RESP, when it is OFF, machine can not
monitor REST, and no parameter area of RESP in main interface.
NIBP: ON/OFF. When it is ON, machine can monitor NIBP, when it is OFF, machine can not
monitor NIBP, and no parameter area of NIBP in main interface.
IBP<1,2>(optional): ON/OFF. When it is ON, machine can monitor IBP<1,2>, when it is OFF,
machine can not monitor IBP<1,2>, and no parameter area of NIBP in main interface.
CO2(optional): ON/OFF. When it is ON, machine can monitor CO2, when it is OFF, machine
can not monitor CO2, and no parameter area of CO2 in main interface.
AG(optional): ON/OFF. When it is ON, machine can monitor AG, when it is OFF, machine
can not monitor AG , and no parameter area of NIBP in main interface.
Caution
When CO2 module is set to "On", RESP option is the system default to "Off".
Caution
When the temperature module manufacturers setting temperature module is locked,
this menu "TEMP" is not option.
Caution