105
levels
NS alarming limit
error
Function/
safety
failure
High
Stop using the NIBP module alarming
function and inform biochemical engineers
or our maintenance team
NM alarming limit
error
Function/
safety
failure
High
Stop using the NIBP module alarming
function and inform biochemical engineers
or our maintenance team
ND alarming limit
error
Function/
safety
failure
High
Stop using the NIBP module alarming
function and inform biochemical engineers
or our maintenance team
Technical alarms 2 (displayed in the reminder area below NIBP values):
Reminders Causes
Alarming
levels
Solution
NIBP self-detection
error
Sensor or other
hardware failure
High
Stop using the NIBP measurement
function and inform biochemical
engineers or our maintenance team
NIBP
communication
error
Communication
failure with
NIBP
measurement
module
High
If the failure continues, stop using the
NIBP measurement function and inform
biochemical engineers or our maintenance
team
Cuff loose or out of
connection
Cuff isn’t
properly
wrapped or no
cuff
Low Wrap the cuff properly
Gas leakage with
cuff pumping pipe
Damage with
cuff, pipe or
connector
Low
Check and change the part with leakage,
and if necessary also inform biochemical
engineers or our maintenance team
Air pressure error
Fail to obtain
stable pressure
value, e.g.,
because of pipe
wrapping
Low
Check whether there is pipe wrapping; if
the problem continuous, inform
biochemical engineers or our maintenance
team
Signal too weak
Cuff too loose
or patient pulse
too weak
Low
Apply other methods to measure blood
pressure
Pressure out of
scope
Measurement
scope exceeding
specified upper
limit
High
Reset the NIBP measurement module;
and if the problem continuous, stop using
the NIBP measurement module and
inform biochemical engineers or our
maintenance team
Arm movement Big signal noise Low Make sure the patient is silent, without