106
or irregular PR
due to arm
movement
movement
Overpressure
protection
Pressure
exceeding
specified upper
limit
High
Re-measure. If the problem continuous,
stop using the NIBP measurement module
and inform biochemical engineers or our
maintenance team
Signal saturation
Significant
movement
Low Stop the patient from movement
Pump leakage
Leakage during
the leakage test
Low
Check and change the part(s) with
leakage, and if necessary, inform
biochemical engineers or our maintenance
team
NIBP system failure
Blood pressure
pump system
running failure
High
Stop using the NIBP measurement
function and inform biochemical
engineers or our maintenance team
Wrong cuff type
Current cuff
type not suitable
for patient
Low Select proper cuff
Measurement time
out
Measurement
time over 120
seconds (for
adults) or 90
seconds (for
newborns)
High
Re-measure or apply other measurement
methods
NIBP reset error
Module reset
improperly
High Re-use the reset function
Measurement error
During
measurement
the system can
not execute
measurement
analysis or
calculation
High
Check the cuff and re-measure while
making sure the patient is silent during
monitoring
Reminders (displayed in the reminder area below NIBP values):
Reminders Causes Alarming levels
Manual measurement… Manual measurement in process
No alarm
Continuous measurement… Continuous measurement in process
Automatic measurement… Automatic measurement in process
Please press START After you select the measurement time
interval