122
d) Connect the interface cable of the ISA analyzer to the CO
2
interface of the CO
2
plug-in
module or the minihost.
e) Enter into the conventional screen of the host monitoring equipment, select [Exchange
Waveform] to call out the “CO
2
” waveform and parameters which you want to monitor, such
as [CO
2
] (this step can be skipped if the screen has already displayed the “CO
2
” waveform
and parameters).
f) When the CO
2
module is connected to the monitor, its module working mode is in the
“Measurement” state; however, in order to make sure that it is in the correct working state,
please do enter into the [CO
2
Settings] menu to set its “Working Mode” to the [Measurement]
mode.
g) Set [CO
2
Switch] to [On].
h) Set appropriate [Oxygen Compensation], [Laughing Gas Compensation].
i) To connect the outlet of the sample gas to the discharge system, or to make the gas to flow
back to the patient’s circuit.
j) If it is green LED indication, ISA Analyzer is available.
k) To carry out inspection before use according to the statement in the “Inspection Prior to Use
(2)”.
l) If the inspection is normal, start to monitor the Anaesthetic Gas.
16.4.2.2 Check before use
Before connecting the Nomoline sampling pipe to the breathing circuit, carry out the following
steps:
a) Connect the sampling tube to the gas entrance interface (LEGI) of the ISA CO
2
module.
b) Check whether the green light of LEGI is steadily on or notThe indication system is
normal. .
c) Exhale to the sampling tube, check if a valid CO
2
waveform and value are displayed on the
host monitoring equipment.
d) Use the finger tip to block up the sampling pipe, and hold on for 10 minutes.
e) Examine whether there is obstruction warning and if the LEGI shows a red flashing light.
f) Under proper circumstancesCarry out enclosure check on the patient’s circuit that is linked