100
Fig 13-4 NIBP Setup
ALM ON/OFF(Alarm switch): Where “ON” is selected, the alarm prompt and
saving will proceed when the pressure is alarmed; where “OFF” is selected,
alarming will not happen, but
will be prompted beside NIBP in the screen
parameter area.
ALM REC(Alarm record): Users can select “On” to output through recorder when
blood pressure alarms happen
Measure mode:big or small
INTERVAL: MANUAL and auto(1min, 2min , 3min, 4min,
5min ,10min ,15min ,30 min ,60min ,90min ,120min ,180 min ,240 min ,480min)
DISP COLOR(Display color): green, cyan, red, yellow, white, blue, violet.
NIBP SYS / NIBP MEA / NIBP DIA ALM STEUP: set the valve of NS ,NM ,ND
high limit ,middle limit ,low limit.
NS, ND, NM upper/lower limits
Big:
NS upper limi:42-270 mmHg lower limit:40-268 mmHg
ND upper limi:12-210 mmHg lower limit:10-208 mmHg
NM upper limi:22-230 mmHg lower limit:20-228 mmHg
Small:
NS upper limi:42-200 mmHg lower limit:40-198 mmHg
ND upper limi:12-150 mmHg lower limit:10-148 mmHg