z The ISA sidestream gas analyzer is designed to be used by authorized or trained
medical personnel.
z Only Nomoline sampling tubes produced by PHASEIN can be used.
z The ISA sidestream gas analyzer shall not be used in an inflammable anesthetic
gas.
z You should earnestly neaten the sampling tube in order to reduce the risk of it
wrapping or reining the patient.
z Do not repeatedly use a disposable sampling tube.
z Do not lift the ISA/host equipment by grasping the sampling tube; otherwise it may
break away from the ISA/host equipment, which may result in that the ISA/host
equipment falls onto the patient.
z Used disposable sampling tubes should be disposed according to local medical
waste stipulations.
z Do not use a sampling tube configured for a big animal on a small animal;
otherwise the invalid cavity in the patient circuit will increase.
z Do not use a sampling tube configured for a small animal on a big animal;
otherwise it will result in a too big flowing resistance.
z Do not use the ISA sidestream gas analyzer together with a quantitative spraying
agent or spray; otherwise it may result in the clogging of the germ filter.
z Check if the flowing speed of the gas sample is too high for the given patient type.
z Since successful zeroing requires that the gas analyzer exists in the ambient air
(21% oxygen and 0% CO2), you should make sure that the ISA is placed at a well
ventilated position. Before and after executing the zeroing procedure, avoid
breathing in the vicinity of the ISA sidestream gas analyzer.
z The Nomoline sampling tube and its interface are not germ free devices. In order
to prevent the sampling tube from causing damages, please never carry out high
pressure sterilization on any part of the sampling tube.
z Never disinfect the ISA sidestream gas analyzer or soak it into a liquid.
z Mobile and radio frequency communication equipment will affect measurement.
Make sure that the ISA gas analyzer is used in the electromagnetic environment
designated in this operating instruction manual.
z The ISA gas analyzer can only be used as a piece of auxiliary equipment for patent