—Pulse Rate(unit:bpm)
IBP (optional)
— Invisive Blood Pressure
(From left to right) Systolic, Diastolic, Mean (⑦, Unit: mmHg or kPa)
TEMP
—TEMP
—Temperature(⑧,unit: or℃℉)
CO2 (optional)
—end-tidal CO2 (⑨, Unit: mmHg or kPa)
—inspiratory CO2 (⑩, Unit: mmHg or kPa)
—Air Way Respiration Rate (AWRR)(
○
11 ,unit:bpm)
RESP
—Respiration Rate(,unit:bpm)
The above monitored results will show in the parameter area.
The parameters are refreshed once per second, but NIBP value, once per measurement.
Users can select the monitoring parameters and the main screen will display the relative content.
Menu Area(④)
1. PATIENT: patient information configuration. Refer to chapter of Patient Informatio for details.
2. ALARM SETUP: set up alarm type
ALARM RECORD TIME: in case of physiological alarm, the system can record the
information before and after the alarm time. The system can give three kinds of time, i.e., 8
seconds, 16 seconds. The seconds in the options are the sum of seconds before and after the
alarm time. For example, 8 seconds mean the information within 4 seconds before the alarm
time and within 4 seconds after such a time.
ALM PAUSE TIME (Alarm pause time):“1 minute”, “2 minutes” or “3 minutes”
ALAM LIMIT (Display alarm limits):”ON” and”OFF”
ALM LATER :Disabled ,5s,10s,15s,20s.
Select "Disable", when an alarm occurs, the
instrument that occurs immediately alarm; When Select "five seconds", alarm occurs only
when the parameters continue to exceed the the alarm upper or lower limit of 5 seconds
instrument, other settings and so on. The purpose of this feature is to reduce the instant false
alarms, allow doctors to more accurately grasp the situation
3. SURVEY SETUP: ECG, RESP, SpO2, NIBP and TEMP setup, refer to each chapter for details
4. SCREEN CHANGE: STANDARD, LIST FACE, TREND SCREEN, oxyCRG SCREEN, BIG
FONT, AND 7CH MULTI-LEADS DISPLAY. For details,please refer to chapters for each