9.6.1 Lowering the warm-up temperature
Use the warmup setpoint temp menu item and decrease the warmup set point to the
desired temperature. The minimum temperature allowed is 0 °C.
9.6.2 Disabling the warm-up cycle
It may be required for the pump to skip the warmup cycle and go directly onprocess
once it has reached full speed. There are two ways to achieve this:
Enable COLD ONPROCESS
Adjust the WARMUP Setpoint temp set point to below ambient temperature.
9.6.3 Increasing the warm-up temperature
Use the Warmup setpoint temp menu item and increase the warmup set point to the
desired temperature to suit your applicaon.
Note:
The maximum temperature allowed for warmup setpoint temperature is above the
alarm limit for the pump so if the temperature is set too high the pump will never
warm
up. Changing the warm
up temperature does not change the pump's water
cooling operaon. This is xed by us and can only be changed by a downloadable
conguraon.
9.6.4 Stop the pump going on-process when it is warm
The pump can be congured to automacally go into the Green mode/standby mode
when it is warm instead of going onprocess. To do this, go to auto Green Mode menu
and disable auto Green Mode.
Note:
If the pump is controlled by a tool interface then the pump will follow the state of the
on
process line.
9.6.5 Stop the pump automacally warming up when in Green mode/standby
mode
If the pump is running in the Green mode/standby mode and its internal temperature
drops below the warmup set point then by default the pump will automacally perform
a warmup cycle. This could result in the pump speeding up and slowing down over a
period of me.
If the applicaon does not require the pump to remain above the warmup set point
while in the Green mode/standby mode then automac rewarming can be disabled
using the Warmup auto rewarm menu item.
9.6.6 Turn the inlet purge on during warm-up
A pump can be made to warmup more quickly by adding inlet purge during the warm
up cycle. This is enabled using the Inlet purge menu item.
Page 136
M58800880_H - Pump display terminal

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards GXS Series and is the answer not in the manual?

Edwards GXS Series Specifications

General IconGeneral
BrandEdwards
ModelGXS Series
CategoryWater Pump
LanguageEnglish

Related product manuals