This translates to:
alarm 1 in bit 2 ‑ PWM_TRIP
alarm 8 in bit 4 ‑ OVERT
warning 2 in bit 6 - HIGHTM.
This means the motor is overheated and switched the output o.
10.5 Other problems
10.5.1 Pump controller communicaons
The pump controller contains two processors that perform dierent funcons. Under
normal operaon, the two processors communicate with each other internally while
they perform their funcons.
Under fault condions, it is possible that the two processors stop communicang with
each other. In this circumstance:
▪ The pump may connue to run normally, with gas valves and gate valves kept in
the same state.
▪
Depending which device is 'in control' of the system, control might be released (for
example PDT).
▪ The front panel controls can be used to 'take control' and stop the pump but this
will only be a simple stop, it will not be possible to use any of the shutdown
sequences such as Smart Stop.
▪ If the front panel controls or the EMS buon are used to stop the pump, the
warning LEDs on the pump will ash and the alarm LED will illuminate. The inlet
isolaon valve will shut.
▪
Stop requests from all the devices apart from the front panel controls will be
ignored.
▪ Any requests to start the pump will be rejected.
For systems ed with the MCM MicroTIM:
▪
If the pump is stopped when communicaon is lost then 'alarm present' is set on
the alarm line.
▪ If the pump is running when communicaon is lost then the outputs to the
MicroTIM are unaected.
Page 173
M58800880_H - Fault nding