10. Depending on the pump size, it typically takes around 4‑7 minutes to draw through
the 5 litres of the solvent soluon (larger pumps draw the liquid through more
quickly than the smaller pumps). Once the full 5 litres of the solvent soluon have
been drawn through, put the free end of the solvent sucon pipe right down to the
boom of the container holding the 2.5 litres of water. The water will ush
through any solvent soluon remaining in the pump mechanism.
11. The pump typically takes another 2‑4 minutes to draw through the water. Once all
the water has been drawn into the pump, the DP clean process connues with a
purge cycle, drawing air or purge gas into the pump to dry out the mechanism for
the rest of the clean me.
Notes:
1. If monitoring the state of the solvent that exits the pump to determine whether
the pump is clean, take special care because the solvent soluon inially exits the
pump as steam and then as hot liquid with typical temperature of 75 °C.
2. The exhausted liquid will never run clear. The exhausted liquid from a clean
pump has a similar viscosity to water (not thick and viscous) and is a light rust
colour when seen exing the exhaust.
3. Depending on the applicaon, the solvent ush process may need to be
repeated. Contact us for advice.
4. If the pump takes an abnormally long me to draw in uids, there may be a
problem with the high ow purge and solvent ush kit. A blocked orice or leaks in
the solvent sucon pipe or pneumac gas line are the most likely causes.
Invesgate the problem and recfy before connuing to use the solvent ush
procedure. If in doubt, contact us.
Once the DP clean me has elapsed, the dry pumping system will return to green mode/
standby mode sengs. The pump can then be put back on‑process.
To stop the solvent ush process while it is running, enter the DP Clean menu again
using the PDT and select 'DP Clean o'. The pneumac valve will shut, stopping the ow
of solvent into the pump. If DP Clean has been congured to run an inlet purge cycle
aer the clean cycle, sending a 'DP Clean o' command will iniate the purge cycle. To
stop the purge cycle as well, the PDT must be used to select 'DP Clean o' for a second
me.
The DP clean procedure can be congured to suit the user's applicaon. Refer to How to
set up the DP clean on page 141.
4.11 Pump speed control and PID
For some applicaons it is important to be able to adjust the speed of the pump while it
is on‑process, either to a set speed or to control the speed connuously to maintain a
process pressure. The dry pumping system can run both the dry pump and booster at
various speeds while on‑process and it is possible to congure these speeds by a
number of dierent methods. Refer to How to control the pump speed on page 144 for
more informaon about how to set up speed control opons.
The dry pumping system has a built‑in Proporonal-Integral-Derivave (PID) controller.
When the PID funcon is enabled and the pump is on‑process, the PID controller
Page 65
M58800880_H - Operaon