3.12 Install addional safety equipment
WARNING:
If the control system needs to know the total ow rate of purge gas to the dry
pumping system for safety reasons, install suitable measurement equipment in the
purge gas supply pipeline.
WARNING:
If the gas purges are used to dilute dangerous gases to a safe level, ensure that the
dry pumping system shuts down if the purge gas supply to the dry pumping system
fails.
If the total ow rate of purge gas to the dry pumping system is required to be known for
safety reasons, suitable measurement equipment should be ed in the purge gas
supply pipeline. If ng a rotameter, ensure that it is suitable for use with the purge gas
and that it is correctly calibrated.
Ensure that the installaon is congured so that it remains safe if there is a failure of the
purge gas supply to the dry pumping system.
If an alarm condion is detected, the dry pumping system will shut down automacally.
Ensure that the installaon remains safe if the dry pumping system shuts down
automacally.
3.13 Purge gas set up
There are two types of gas module used on dry pumping systems. Refer to Purge data on
page 18. The gas module conguraon may be adjusted to suit process demands.
Contact our service personnel for instrucon on how to access the gas valves menu (by
the PDT) and for the processspecic recommendaons.
For dry pumping systems using the mediumduty gas module, there is an adjustable
needle valve ed to the purge pipe which enables the adjustment of the level of gas
ballast to suit the applicaon.
Use the following procedure to adjust the gas ballast:
1.
Remove the le hand side panel of the dry pumping system.
2.
Idenfy the adjustable needle valve on the purge pipe (contact us for guidance).
3.
Adjust the needle valve unl the required ow is met. Refer to Purge data on page
18 for the purge gas ows.
4.
Ret the le side panel.
For GXS450 and GXS750 dry pumping systems with mediumduty gas module, there is
an addional electronically controlled gas ballast available for harsh applicaons. By
default, these dry pumping systems are supplied with this addional gas ballast switched
o but it can be enabled using the PDT.
Use the following procedure to enable the addional gas ballast:
1.
Press the SETUP buon on the PDT to enter the SETUP menu.
2.
Scroll using the UP/DOWN keys and select the FIT ACCESSORY menu.
Page 50
M58800880_H - Installaon

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards GXS Series and is the answer not in the manual?

Edwards GXS Series Specifications

General IconGeneral
BrandEdwards
ModelGXS Series
CategoryWater Pump
LanguageEnglish

Related product manuals