Evacuang the chambers back‑lled with Argon
▪ For small chambers less than 2 m
3
(70 c) or if chamber pressure reaches 20 mbar
in under 2 minutes, the chamber evacuaon is short enough not to cause a
problem when pumping at full speed.
▪ For evacuang chambers between 2 m
3
(70 c) and 25 m
3
(885 c), the pump
speed should be set at 80 Hz unl the pressure is below 20 mbar.
▪ For evacuang chambers larger than 25 m
3
(885 c), contact us for more
informaon.
Note:
For prolonged operaon (greater than 45 minutes, above 20 mbar and/or the
chambers in excess of 25 m
3
(885 c)), an applicaon review must be done to
determine the actual requirement. Contact us for more informaon.
Sha seal purge
Nitrogen or the Clean Dry Air (CDA) purge is preferred as it gives a maximum headroom
on the crical clearances. Argon can be used as sha seal purge at 80 Hz and 110 Hz,
following the above guidelines for connuous operaon and pump down of chambers.
4.5 Manual shut down
WARNING:
If the dry pumping system is shut down and not isolated from the electrical supply,
do not disconnect the PDT or release the control from the PDT or front panel. This
may result in the dry pumping system being started by another module.
WARNING:
Do not remove the inlet connecons unl the pump has been allowed to stop
rotang and the power has been isolated. The pump can take up to three minutes to
completely stop.
CAUTION:
If the pump is stopped without the gas purge cycle on processes that have
condensable or solid by‑products then the pump may not restart.
CAUTION:
For applicaons pumping liquids or condensible gases: If the pump is switched o
straight aer the process then the liquids may remain trapped in the pump. These
liquids can corrode the pump mechanism while the dry pumping system is switched
o for extended periods and may prevent the pump from starng again. For these
applicaons, we recommend carrying out a Smart Stop with Stop Time set to the
maximum 3600 seconds. Once the pump has stopped, seal the inlet and exhaust to
prevent migraon of liquid/condensable back into the pump mechanism.
The dry pumping system has two manual shut down modes: Fast and Auto.
Page 59
M58800880_H - Operaon