100 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x 14 Data Logger and Transmission of Data
14.1.2 Setting the Storage Rate
The storage rate is the frequency at which the measured values are transmitted or sto-
red. The storage rate will be set separately for each measuring channel.
If the storage rate is not set, the storage rate previously selected will be used.
The storage interval should be at least equal to the number of activated measuring chan-
nels, e.g. the storage interval of a channel should be min. 2 s if 2 measuring channels are
activated, min. 4 s are recommended.
14.1.3 Settings for the Data Logger
Select program branch Special Funct.\SYSTEM settings\Storing. It contains
the following menu items:
• ringbuffer
• storage mode
• storing of the totalizer values
• storing of the signal amplitude
• storing of the sound speed
• acoustic signal during the storing
Press ENTER until the menu item Store Meas.Data is
displayed.
Select yes to activate the data logger. Press ENTER.
Select a storage rate or EXTRA. Press ENTER.
This display will only be indicated if Store Meas.Data
and/or Serial Output are activated.
If EXTRA has been selected, enter the storage rate. Press
ENTER.
Storage Rate ↕
Once per 10 sec.