14 Data Logger and Transmission of Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 101
Ringbuffer
The setting of ringbuffer affects the storing of measured values as soon as the data me-
mory is full:
• If the ringbuffer is activated, the available data logger will be halved. The oldest mea-
sured values will be overwritten. Only the data logger memory that was free during the
activation will be used by the ringbuffer. If more data logger memory is necessary,
measured values in the data logger should be deleted previously.
• If the ringbuffer is deactivated, the storing of measured values will be stopped.
Storage Mode
Storing of the Totalizers
see section 13.2, Selection of the Totalizers for Storing
Storing of the Signal Amplitude
Select the behavior of the ringbuffer. Press ENTER.
If on is selected, the available data logger memory will be
halved. The oldest measured values will be overwritten.
If off is selected, the storing of measured values will be
stopped.
Select the storage mode. Press ENTER.
If sample is selected, the displayed measured value will
be used for storing and online transmission of data.
If average is selected, the average of all values
measured during a storage interval will be used for storing
and online transmission of data.
Note! The storage mode does not affect the outputs.
Note! Storage mode = average
The average of the physical quantity and other physical quantities
assigned to the measuring channel, e.g. the measured temperature,
will be calculated.
If the storage rate < 5 s (see section 14.1.2) is selected, sample will
be used.
If no average could be calculated over the complete storage interval,
the value will be marked as invalid. The ASCII file will contain "???"
instead of invalid average values of the physical quantity and
"?UNDEF" instead of invalid temperatures.
If on is selected and the data logger is activated, the amp-
litude of the measured signal will be stored together with
the measured values. Press ENTER.
Storage mode
>SAMPLE< average