102 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x 14 Data Logger and Transmission of Data
Storing of the Sound Speed of the Medium
Storing of the Diagnostic Values
Acoustic Signal
Per default, an acoustic signal will be emitted every time a measured value is stored or
transmitted to a PC or printer. The signal can be deactivated in Special Funct.\
SYSTEM settings\Storing\Beep on storage.
14.1.4 Measurement with Activated Data Logger
• Start the measurement.
If Output Options\Store Meas.Data has been activated and Special Funct.\
SYSTEM settings\Ringbuffer is deactivated, this error message will be displayed as
soon as the data logger is full.
If no other output (transmission of data, outputs) has been activated, the measurement
will be stopped.
If another output has been activated, the measurement will be continued. Only the sto-
ring of the measured values will be stopped.
If on is selected and the data logger is activated, the sound
speed of the medium will be stored together with the
measured values. Press ENTER.
If on is selected and the data logger is activated, the dia-
gnostic values will be stored together with the measured
values. Press ENTER.
Select off to deactivate the acoustic signal, on to activate
it. Press ENTER.
Enter the measuring point number. Press ENTER.
If arrows are displayed in the lower line on the right, ASCII
text can be entered. If digits are displayed, only digits,
point and hyphen can be entered.
For the setting of the input mode see section 17.2.3.
Press ENTER.
The error message will be displayed periodically.
Store diagnostic
off >ON<
A:Meas.Point No.:
xxx (↑↓← →)