Versiv Series Cabling Certification Product Family
Technical Reference Handbook
102
Figure 38. Wire Map Examples: Open Wire and Open Shield
GPU60.PNG
GPU61.PNG
Open
Wire 4 is open 36 m from the tester
and 14 m from the remote.
Open Shield
The shield is open 3.6 m from the
tester. The wire map always shows
the shield if you select a shielded
cable type (such as S/FTP) on the
TEST SETUP screen and you turn on
the Shield Test. If you select an
unshielded cable type or turn off
the Shield Test, the wire map shows
the shield only if the shield has
continuity.