601
Chapter 21: Specifications
Notes
Specifications can change without notice. For the latest
specifications, go to the Fluke Networks website.
Specifications apply at 23
o
C (73
o
F), unless otherwise
noted.
Environmental and Regulatory Specifications
Operating temperature
1, 2, 3
With DSX CableAnalyzer modules:
32 °F to 113 °F (0
o
C to 45
o
C)
With CertiFiber Pro or OptiFiber Pro modules:
14 °F to 113 °F (-10
o
C to +45
o
C)
Storage temperature
4
With battery: 14 °F to 140 °F (-10
o
C to +60
o
C)
Without battery: -22 °F to +140 °F (-30
o
C to
+60
o
C)
Operating relative humidity
(% RH without condensation)
0 % to 90 %, 32 °F to 95 °F (0
o
C to 35
o
C)
0 % to 70 %, 95 °F to 113 °F (35
o
C to 45
o
C)
Uncontrolled: <0
o
C
Vibration Random, 2 g, 5 Hz-500 Hz
Shock 1 m drop test with and without module and
adapter
1. Using battery power. With ac power: 0
o
C to 45
o
C.
2. With the OptiFiber Pro OTDR module: Real Time Trace function used for no more
than 5 minutes in a 15-minute period. Maximum ambient temperature is 35
o
C for
continuous use of the Real Time Trace function.
3. The battery will not charge if its temperature is outside the range of 32 °F to 104 °F
(0 °C to 40 °C).
4. Do not keep the battery at temperatures below -20
o
C (-4
o
F)
or above +50
o
C
(+122
o
F) for periods longer than one week. If you do, the battery capacity can
decrease.
-continued-