Chapter 11: The OptiFiber Pro OTDR Module
OTDR Results
359
reference for the calculation of the bend’s loss. If you need to
see the bend’s loss at each wavelength, do the test again with
Macrobend Detection turned off.
THRESHOLD: If the difference between the loss
measurements for the wavelengths is more than the
threshold, the tester identifies the event as a bend. The
default is 0.50 dB.
MARGIN: The THRESHOLD value minus the delta value ().
Note
Macrobend detection is available only for
singlemode fibers.
About Measurements that Show “<“ and “>”
In some situations, measurements show with a "<" or ">" (less
than or greater than) symbol. This occurs if the fiber is too long,
the signal is weak, events are very close together, or a
measurement is beyond the range of the tester.