Versiv Series Cabling Certification Product Family
Technical Reference Handbook
290
Bi-directional Tests
Do bi-directional tests when they are required by the
manufacturer or by your customer.
The tester can automatically do bi-directional tests in Smart
Remote and Loopback modes. To get bi-directional results in Far
End Source mode, do a test from each end of the fiber.
In Smart Remote mode, the tester saves bi-directional results in
two records. Each record contains the results for one fiber for
both directions.
To do a bi-directional test
Note
When you make connections for a bi-directional
test in Smart Remote mode, connect the fiber that
has the Input Fiber ID to the main tester’s OUTPUT
port, and the fiber that has the Output Fiber ID to
the main tester’s INPUT port. Halfway through the
test, you will switch the connections.
1 Set the reference and measure the loss of the TRCs for the
mode you will use:
Smart Remote mode: See “Step 1: Set the Reference in
Smart Remote Mode” on page 260 and “Step 2: Measure
the Loss of the Test Reference Cords You Will Add” on
page 264
Loopback Mode: See“Step 1: Set the Reference in
Loopback Mode” on page 274 and “Step 2: Measure the
Loss of the Test Reference Cord You Will Add” on
page 276.
2 Connect to the link as shown on the left side of Figure 123.
3 On the home screen, tap the test setup panel, make sure the
correct test is selected on the CHANGE TEST screen, then tap
EDIT.