Connector icons show the connections between the ends of the
link. If you enter 7 or more for the Total Connections setting, a
number inside of a connector icon shows the number of
connectors between the ends of the link. For example, if the
Total Connections setting is 7, a connector icon shows the
number 5 ( )
The round icon shows the number of splices in each path of the
link.
To see help for the screen, tap .
To save your settings, tap DONE.
About 1 Jumper Reference Connections
The reference and test connections shown in this manual give 1
jumper results. 1 jumper results include the loss of the fiber plus
the loss of the connections at both ends of the link. This is the
best method for tests on premises fiber installations. Premises
installations typically use patch cords at both ends of the link, and
connector loss is a large part of the total loss.
If you do not have the correct connector adapters, see Appendix C
for other connections that give 1 jumper results.
For descriptions of the 2 and 3 jumper reference connections, see
Appendix C.
WCaution
Most cable manufacturers will give you a warranty
on a fiber installation only if you use a 1 Jumper
reference when you certify the installation.
Note
If you use a 2 Jumper reference, the Wizard for
the reference procedure does not show steps for
the TRC verification. To save test results for the
TRCs, do the tests manually.