Fluke DSX-8000 - Step 2: Measure the Loss of the Test Reference Cords You will Add

Fluke DSX-8000
700 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Versiv Series Cabling Certification Product Family
Technical Reference Handbook
264
Step 2: Measure the Loss of the Test Reference Cords You
Will Add
WCaution
If you disconnected a test reference cord from the
output of the tester or remote, you must set the
reference again to make sure your measurements
are reliable.
2-1 On the SET REFERENCE screen, when the reference procedure
is completed, tap NEXT.
2-2 Disconnect the test reference cords from the INPUT ports on
the tester and remote, then use test reference cords and
adapters to make the connections to verify the TRCs, as shown
on the screen and in Figure 111.
2-3 Tap TRC VERIFICATION. The tester measures and saves the loss
of the test reference cords you added. The IDs for these results
start with “TRC”, show the date and time of the test, and have
an for the test result.
The tester shows a warning if the loss of a TRC is more than
these limits:
Maximum loss for multimode TRCs: 0.15 dB
Maximum loss for singlemode TRCs: 0.25 dB
If the tester shows a warning, clean and inspect the
connectors on the TRCs in the path that has too much loss,
make sure the cords are straight as shown in Figure 111,
then do the TRC verification again.