Versiv Series Cabling Certification Product Family
Technical Reference Handbook
612
Dynamic
Accuracy
NEXT
± 0.75 dB
Dynamic
Accuracy
ACR-F
± 1.0 dB (FEXT dynamic accuracy is tested to ± 0.75 dB)
Source/load
return loss
1to 300MHz:
20 – 12.5log(f/100),
20 dB maximum
300 to 600 MHz:
14 dB
600 to 1000 MHz:
21-9 log(f/100)
1 to 300 MHz: 18 – 12.5 log(f/100),
20 dB maximum
300 to 600 MHz: 12 dB
600 to 1000 MHz: 19-9log(f/100)
Random
Noise Floor
95 dB 90 dB
Residual NEXT
90 – 20 log(f/100)
(measured to 85 dB
maximum)
85 - 20 log(f/100)
(measured to 85 dB
maximum)
1 to 600 MHz:
82.3 - 20 log(f/100)
600 to 1000 MHz:
97.9 - 40 log(f/100)
(measured to 85 dB
maximum)
Residual FEXT
80 - 20 log(f/100)
(measured to 85 dB
maximum)
75 - 20 log(f/100)
(measured to 85 dB
maximum)
60 - 15 log(f/100)
(measured to 85 dB
maximum)
Output Signal
Balance
1 to 600 MHz:
40 - 20 log(f/100)
600 to 1000 MHz:
24.4 dB
(measured to 60 dB
maximum)
1 to 600 MHz: 37 - 20 log(f/100)
600 to 1000 MHz: 21.4 dB
(measured to 60 dB maximum)
DSX-5000 Level V Accuracy Performance Parameters
per IEC Guidelines (continued)
Parameter Baseline Field Tester Field Tester with
Level V Permanent
Link Adapter
Field Tester with
Level V Channel
Adapter