EasyManua.ls Logo

Fluke DSX-8000 - Page 664

Fluke DSX-8000
700 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Versiv Series Cabling Certification Product Family
Technical Reference Handbook
628
OptiFiber Pro HDR OTDR Module Specifications (cont.)
Reflectance
Accuracy
13
2 dB typical maximum
Sample Resolution
14
3 cm to 2 m
Sampling Points Up to 129,000
Pulse Widths
(nominal)
5, 10, 30, 50, 100, 300, 500, 1000, 3000, 5000, 10000,
20000 ns
Loss Threshold
Setting
0.01 dB to 1.5 dB
Adjustable in 0.01 dB increments
End Threshold Setting 1.0 dB to 30.0 dB
Adjustable in 0.1 dB increments
Units of Measure m, ft
Test times per wavelength
Auto OTDR 5 seconds/wavelength (typical)
Auto PON OTDR 10 seconds/wavelength (typical)
Quick Test 3 seconds/wavelength (typical)
Best Resolution 5 to 180 seconds/wavelength
FaultMap 10 seconds typical; 180 seconds maximum
Manual 3, 5, 10, 20, 40, 60, 90, 120, 180 seconds/wavelength
Laser classification Class I CDRH, Complies with EN 60825-2 3rd Edition
Real Time Trace
Refresh Rate
0.3 seconds (typical)
Minimum Power for
Live Fiber Detection
-40 dBm at 1310 nm
Maximum Live Fiber
Input Power
+10 dBm continuous wave or data
13. Applies along the trace backscatter within the distance range in which the OTDR can
find a 0.1 dB non-reflective event.
14. Does not apply to the Real Time Trace.