GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-31
RT_ConnectionMngr::vAssembleRTSet: no
RT_TissueFilterBlock found for TISSUE3D Blocks
restart the system
RT_ECG_Block::bStart ECGInterface failed!! restart the system
RT_MColorFilterBlock::RT_MColorFilterBlock: Memory
allocation failed
restart the system
RT_MColorFilterBlock::vCheckIQDataSizeAndUpdateBuffer:
Memory allocation failed
restart the system
RT_MColorFilterBlock::vDebugDrawIQDataCurve: Memory
allocation failed
restart the system
RT_MotionMBlock::execute caused exception:: Addr from
DMA= xx, ReplayBuffAddr= xx, (Line+Header)Size= xx,
blockLength= xx, value i= xx!!
restart the system
RT_PWorCW_Block::replay buffer size smaller than expected restart the system
RT_PWorCW_Block::execute caused exception:: Addr from
DMA= xx, ReplayBuffAddr= xx, (Line+Header)Size= xx,
blockLength= xx, line index= xx!!
restart the system
RT_PWorCW_Block::execute: Error, length of DMA block 2 [xx]
is not aligned to linesize!!
restart the system
Start Bandwidth too small restart the system
Start ET too small restart the system
Start frequency too small restart the system
StoragePool: Not sufficient replay storage for B & CF Mode
available
restart the system
StoragePool: Not sufficient replay storage for ECG available restart the system
StoragePool: Not sufficient replay storage for M Mode available restart the system
StoragePool: Not sufficient replay storage for MC or PW Mode
available
restart the system
System detected severe error, please call technical support. restart the system, call technical support
System detected severe error. Some components like Touch
Panel server may not be registered. Please register Touch
Panel server and restart.
restart the system, call technical support
The Database UserPrograms Corrupted restart the system
The Date format not stored properly close registry, restart, try again
The Handle Unregistered close registry, restart, try again
The Registry not closed restart the system
The System detected a problem with your harddisk. Please do not reboot
or shut down the machine!
Please contact your service department as soon as possible.
if possible,
• save “Full Backup” (see: Section 4-5-3 on page 4-35) to
“DVD/CD”, “Network”, or “Other drive” (including images)
• if not currently done, save all images via Sonoview
(see: Section 4-5-6 on page 4-40) ,
• call technical support
Error Messages Actions