GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-30 Section 3-5 - Connection of Auxiliary Devices
3-5-2-1 Attaching the Voluson i / Voluson e to the Dock Cart (cont’d)
6.) Plug the power cable and the USB-cable into the DVD-drive; the other end of the USB cable plug
to an USB port on the Voluson i / Voluson e.
7.) Mount and connect other peripherals as described in Section 3-5-2-5 and Section 3-5-2-6.
8.) Connect the main power cable to the Power Supply of the Dock Cart; the other end connect to a
hospital-grade main power outlet with the proper rated voltage.
9.) Press the Circuit breaker on the Dock Cart. To start up the system, press the On/Off button
(see: Figure 3-87) on the control panel of the Voluson i / Voluson e system (for ~3 seconds).
Figure 3-29 Peripherals mounted and secured / Power Supply of Dock Cart / USB-port on Dock Cart
Tray recorder for
DVD+/R/RW Drive
USB port
Upper storage shelf
(e.g., B&W printer)
Lower storage shelf
(e.g., Color printer)
Power Supply
Circuit breaker on Dock Cart
main power cable