IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-24 Section 3-5 - Connection of Auxiliary Devices
NOTE: For more detailed Safety Considerations when connecting auxiliary devices to the Voluson i / Voluson
e system, please review: Chapter 1 - Auxiliary Devices Safety on page 1-10.
!! CAUTION:
Power Supplies for additional equipment MUST comply with IEC 60601-1. Auxiliary equipment
with direct main connection requires galvanic separation of the signal and/or control leads.
!! WARNING:
DO NOT attempt to use different peripherals and accessories other than approved and
provided by GE!
Any unauthorized peripherals may cause system failure or damage!
!! WARNING:
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1 (see: Section 10-7-2 on page 10-13).
!! CAUTION:
Please observe that some printers may not be medical devices! If they are not medical devices,
they have to be located outside of the patient environment (acc. to IEC 60601-1 / UL 60601-1).