1-16 Section 1-3 - Safety Considerations
1-3-9 Lockout/Tagout Requirements (For USA Only)
Follow OSHA Lockout/Tagout requirements to protect service personnel from injuries caused by
unexpected energizing or start-up of equipment during service, repair, or maintenance.
1-3-10 Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that “items what were saturated and/or dripping
with human blood that are now caked with dried blood; or which were used or intended for use in patient
care” are “regulated medical waste” for transportation purpose and must be transported as a hazardous
material.
!! NOTICE:
Energy Control and Power Lockout for Voluson i / Voluson e.
When servicing parts of the system where there is exposure to voltage greater than 30 Volts:
Unplug the system
Maintain control of the system power plug
Remove Battery
There are no test points to verify isolation, you must wait for at least 20 seconds for capacitors to
discharge.
Beware that the Power Supply, Front End Processor and Back End Processor may be energized even
if the power is turned off when the cord is still plugged into the AC Outlet.
UT
Signed
Date
TA G
&
LOCKOUT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson i BT06 and is the answer not in the manual?

GE Voluson i BT06 Specifications

General IconGeneral
TypeUltrasound System
ModelVoluson i BT06
ManufacturerGE Healthcare
Display15-inch LCD
PortabilityPortable
ApplicationObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
Software FeaturesAutomated measurements
WeightApprox. 10 kg
Power Requirements100-240V AC
ConnectivityDICOM, USB

Related product manuals