5-54 Section 5-13 - Service Page
5-13-2-2 Update
5-13-2-2-1 FMI from DVD
By means of the FMI FROM DVD
button, the Systems C:\ image is partly or completely updated.
The System Software parts to be upgraded depend on contents of the used System DVD.
NOTE: During “FMI from DVD” the used system configuration (incl. Full Backup) will be stored on R:\.
The previously used System configuration (before FMI from DVD was performed) is simply restored
by activating the “Rollback” function. Operation see: Section 5-14-1-3 on page 5-58.
5-13-2-2-2 FMI from external device
By means of the FMI FROM EXTERNAL DEVICE
button, the Systems C:\ image is partly or
completely updated.
The System Software parts to be upgraded depend on contents of the used System Media
(DVD, System Stick).
NOTE: During “FMI from external device” the used system configuration (incl. Full Backup) will be stored on
R:\.The previously used System configuration (before FMI from external device was performed) is
simply restored by activating the “Rollback” function. Operation see: Section 5-14-1-3 on page 5-58.
5-13-2-2-3 ASI - Additional Software Installation
Click the ASI
button to install additional software (e.g., Process Logger).
The Software parts to be installed depend on the contents of the System Media
(DVD, System Stick) that is used.
5-13-2-3 TCP/IP Buffersize
The TCP/IP Buffersize selects the amount of buffer memory used for DICOM® transfers (both
directions).
5-13-2-4 Activate / Deactivate Service...
This offers GE technicians the possibility to view the entire customer´s desktop and operation system.
Remote access to the Voluson i / Voluson e system requires permission and customer input to run
diagnostics.
If a GE Service technician requests remote access to your Voluson i / Voluson e scanner, activate the
service platform as described in Section 5-11-2-2 on page 5-46.
5-13-2-5 Common Service Desktop (CSD)
Access to the Common Service Desktop (CSD) by entering security user level and password.
Activate the service platform as described in Section 5-11-2 on page 5-45.
PRECONDITION:
The “Boot priority order” in BIOS has to be 1. IDE 0, 2. USB CDROM.
PRECONDITION:
The “Boot priority order” in BIOS has to be 1. IDE 0, 2. USB HDD.