2-4 Section 2-2 - General Console Requirements
2-2-4 Probe Environmental Requirements
Operation: Ambient temperature 18° to 30° C
Storage: -10° to 50° C
NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Site preparation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
2-2-6 System Specifications
2-2-6-1 Physical Dimensions of Voluson i / Voluson e
The physical dimensions of the Voluson i / Voluson e unit are summarized in Table 2-5.
2-2-6-2 Acoustic Noise Output
max. 45 dB(A)
2-2-6-3 Electrical Specifications
Please refer to Section 2-2-2-1 "Voluson i / Voluson e Power Requirements" on page 2-2.
!! NOTICE:
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
!! CAUTION:
Only one person is required to unpack the Voluson i / Voluson e ultrasound unit.
At least two people must be available to unpack the Cart.
Attempts to move the Cart considerable distances (or on an incline) by one person alone, could
result in personal injury, and/or damage to the system.
Table 2-5 Physical Dimensions of Voluson i / Voluson e
Height Width Depth
Weight
(without peripherals)
59 cm / 2.3 inches 358 mm / 14.2 inches 313 mm / 12.4 inches 5 kg / 11 lbs.