4-38 Section 4-6 - Software Configuration Checks
Section 4-6
Software Configuration Checks
4-6-1 System Setup
Press the UTILITIES key and select the SYSTEM item from the “Utilities” menu - which is shown in the
menu area on the left side of the screen. The System Setup desktop offers different pages to check:
4-6-2 Measure Setup
Press the UTILITIES key and select the MEASURE item from the “Utilities” menu - which is shown in
the menu area on the left side of the screen.
The Measure Setup desktop offers three different pages to check.
Table 4-10 Software Configuration (System Setup) Checks
Step Task Expected Result(s)
1
General: Check Date and Time setting Date and Time are correct
2
General: Check that Location (Clinic Name) is correct Location Name is correct
3
General: Check Language settings desired Language is displayed
4
General: Check EUM Language settings desired EUM Language is displayed
5
General: Check “Display” settings “Display” settings assigned as desired by the customer
6
User Settings: Adjust/check “LCD Brightness” setup displays a complete gray scale
7
User Settings: Check settings of the “Patient ID fields settings assigned as desired by the customer
8
User Setting: Check all the User Settings settings assigned as desired by the customer
9
P1-P2-P3: Check assignment of the [Print] tab settings assigned as desired by the customer
10
P1-P2-P3: Check assignment of the [Save] tab settings assigned as desired by the customer
11
P1-P2-P3: Check assignment of the [Export] tab settings assigned as desired by the customer
12
Option: Check that all options are set up correct D = Demo , I = Inactive , P = Permanent
13
Network:
Check DICOM
®, Sonoview and Network configuration
settings assigned as desired by the customer
!! NOTICE:
Parameters and possible adjustments depend on the selected Application.
To view, add, delete, reorder, edit or when creating a new parameter (in the
MEASURE & CALC page), it
is very important that all items are chosen correctly and that the relevant item is highlighted.
For further information refer to Chapter 17 in the Basic User Manual of Voluson i / Voluson e.
Table 4-11 Measurement Setup Checks
Step Task Expected Result(s)
1
Measure & Calc:
Check all settings for all applications
setting assigned as desired by the customer
2
Application Parameters:
Check all settings for all applications
setting assigned as desired by the customer
3
Global Parameters: Check all settings setting assigned as desired by the customer