Chapter 5 - Components and Functions (Theory) 5-55
5-13-2-6 Delete all Patients
1.) Click the DELETE ALL PATIENTS...
button. Following WARNING message appears on the screen.
5-13-2-7 Export System Data
Select the EXPORT SYSTEM DATA button on the “Service Tools” page to Full Backup the System
State. This includes dump-files and text files, the full Service Database informations about probes,
boards, Software, Options and the Event Log File to the DVD Drive.
Operation see: Section 7-4-2 "Export Log’s and System Data" on page 7-9.
5-13-2-8 Battery Calibration
Select the BATTERY CALIBRATION
button to perform fuel gauge recalibration of the battery.
Operation see: 3-6-4-2-1 "Recalibration by means of Internal Battery Charger" on page 3-77.
5-13-2-9 Keyboard Layout
To change the keyboard layout to different languages.
Operation see: Section 6-9 "Modification of Keyboard Layout" on page 6-10.
NOTE: Reconfigure the layout of the keyboard is only useful by changing the concerned keys also.
see: Section 8-7 "Replacement of Key Caps (by special native language keys)" on page 8-25.
5-13-2-10 BootUp Screen
With this button you can give in your own name. You can see it, when you Bootup the Voluson i /
Voluson e.
5-13-2-11 Monitor Test
Select the MONITOR TEST
button to perform color calibration.
Operation see: Section 6-3 "LCD Display (Monitor) Adjustment" on page 6-2.
5-13-2-12 Printer
Installation of printers or editing their settings is possible without entering the Windows Desktop.
Operation see: 3-7-4 "Printer Installation manually" on page 3-87.
5-13-2-13 Probe EEPROM
NOTE: “Probe EEPROM” is available from software 8.2.2 and software CC200 BT’14 and higher.
Figure 5-30 Warning message
!! WARNING:
If you select the YES button, all patients data, studies, images and measure report data
will be deleted permanently from the hard disk and cannot be recovered!
!! WARNING:
Only accessories explicitly recognized and released by the system manufacturer GE
may be used in connection with the system.
It is recommended to connect and re-program one probe after the other.