3-5-13-1 Direct Connection to the Voluson i / Voluson e system (cont’d)
NOTE: Dip switches are located on the front respectively beneath the video converter, see Figure 3-71 below.
5.) If not already done, set the switches to the correct positions, as shown in Figure 3-72 below.
6.) When all power and signal cables are connected to the system, VCR, and video converter, proceed
as follows (in the order stated):
a.) Press the power ON switch on the VCR.
b.) Press the power ON switch on the video converter.
c.) Turn ON power to the Voluson i / Voluson e system.
7.) Press the Input Select button (see Figure 3-71 on page 3-57) on the video converter three times
in order to select RGB Input as your signal source.
The LED located on the video converter front panel will be illuminated, accordingly.
NOTE: If the VGA cable is disconnected/reconnected while the system is ON, the VGA output is maintained.
NOTE: If system is shut-down to standby or Full-shut-down while the VGA device is connected continuously,
the VGA output will appear after the system is turned ON.
NOTE: If the system is shut-down to Standby or Full-shut-down and the VGA device is disconnected, the VGA
output will not appear after power is turned ON, and the above procedure will have to be repeated.
Figure 3-71 Video Converter Ophit (left image)/Video Converter Canopus (right image)
Figure 3-72 Video Converter Dip Switch Positions
Dip Switches
Power On/Off button
Video Scan Converter Ophit
Dip Switches
Input Select button
Video Scan Converter Canopus
Dip-Switch Settings NTSC: 5 off
Dip-Switch Settings PAL: 5 on
Model Ophit
Model Canopus