IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 3 - Setup Instructions 3-77
3-6-4-2-1 Recalibration by means of Internal Battery Charger
1.) Press the UTILITIES
key and select the SYSTEM from the menu area on the left side of the screen.
2.) Select the
SERVICE page on the screen. The “password window” appears automatically.
3.) Enter the password SHE and click the ACCEPT
button to display the Service Tools window.
4.) Click on the BATTERY CALIBRATION
button.
5.) Confirm the following dialog with YES
.
6.) Follow the steps as described in the “Battery Calibration wizard” until Recalibration is finished.
Figure 3-93 Service Tools window
Figure 3-94 Confirm with YES
Battery
Calibration
Calibration
Battery
with Software 6.x.x and 7.x.x with Software 8.x.x