GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
8-12 Section 8-2 - System Software - Installation/Upgrade Procedure
8-2-1-5 System Software - Installation Procedure (FMI from DVD) (cont’d)
NOTE: If the BT warning dialog (Figure 8-15 below) appears, enter the Permanent key previously recorded
(refer to Figure 8-2: Option page (e.g., Voluson i) on page 8-5) and then click SUBMIT
.
21.)Remove the System DVD from the DVD/CD+(R)W Drive drive.
22.)Reconnect the external devices, install all the printers and adjust the printer settings as described
in Section 3-7 "Printer Installation" on page 3-80.
23.)Check and match system settings in the
P1-P2-P3 page´s [Print], [Save] and [Export] tab.
24.)Confirm that the date and time are set correctly and that the Windows automatic DST feature is off.
Refer to Section Section 6-10 "Daylight Saving Time (DST) - New Dates" on page 6-12.
25.)In case that Application software was below SW 7.0.0 (prior to system upgrade), check and match
Network drive settings (see: Figure 8-3 on page 8-6) according to the printout.
Figure 8-15 Enter Permanent key
enter “Permanent key”
then click [Submit]