GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-17
8-2-2-5 System Software - Installation Procedure (FMI from external device) (cont’d)
In this case, connect the Power supply to charge the battery, afterwards try again.
NOTE: To check capacity of the battery, move the cursor over the Battery Icon in the status area
(see: Figure 5-11 on page 5-31).
10.)Verify that only the DVD drive is connected to the system, then click OK
(Figure 8-20 below).
11.)To start update procedure click YES
(Figure 8-21 below).
12.)The system saves Full Backup in silent mode on R:, then it reboots into LINUX.
13.)A silent “Rollback” image from C:\ is stored on R:\.
After executing all LINUX commands, the system reboots again.
Figure 8-19 battery less charged or Power supply is disconnected
Figure 8-20 Verify that USB devices are disconnected, then click OK
Figure 8-21 Yes - start update procedure