8-68 Section 8-20 - Replacement of the Voluson Station Components
8-20-9 Replacement of the optional Probe Mux Box
8-20-9-1 Manpower
One person, 5 minutes
8-20-9-2 Tools
Phillips screwdriver 1
8-20-9-3 Preparation
1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
2.) Remove the Voluson i / Voluson e from the Tray.
8-20-9-4 Probe Mux Box - Removal Procedure
1.) Disconnect the Probe connector.
2.) Hold the Probe box and screw off the 4 screws.
Figure 8-88 disconnect the Probe connector
Figure 8-89 unscrew the 4 screws