GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-75
8-20-15 Replacement of the Bottom Column Back Cover
8-20-15-1 Manpower
Oner person, 15 minutes
8-20-15-2 Tools
Phillips screwdriver 1
8-20-15-3 Preparations
1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
2.) Remove the Storage Tray Back as described in Section 8-20-3 on page 8-62.
3.) Remove the Top Shelf Additional Bottom Cover as described in Section 8-20-12 on page 8-72 or
remove the optional Probe Mux Bos as described in Section 8-20-9 on page 8-68.
4.) Remove the Power Supply Cover as described in Section 8-20-11 on page 8-71.
5.) Remove the Top Shelf Bottom Rear Cover as described in Section 8-20-13 on page 8-73.
6.) Remove the Top Column Back Cover as described in Section 8-20-14 on page 8-74.
8-20-15-4 Bottom Column Back Cover - Removal Procedure
1.) Screw off the 4 screws and carefully remove the Bottom Column Back Cover.
8-20-15-5 Bottom Column Back Cover - Installation Procedure
1.) Place the new Bottom Column Back Cover and fasten it with the 4 screws.
2.) Mount the Top Column Back Cover as described in Section 8-20-14-5 on page 8-74.
3.) Mount the Power Supply Cover as described in Section 8-20-11-5 on page 8-71.
4.) Mount the Top Shelf Bottom Rear Cover as described in Section 8-20-13-5 on page 8-73.
5.) Mount the Top Shelf Additional Bottom Cover as described in Section 8-20-12-5 on page 8-72 or
mount the optional Probe Mux Box as described in Section 8-20-9-5 on page 8-69.
6.) Mount Storage Tray Back as described in Section 8-20-3-5 on page 8-62.
Figure 8-98 unscrew the 4 screws