6-3
TITRATION MODE
6 TITRATION MODE
6.1 Running a Titration
Before beginning a titration, make sure that the following conditions are met:
At least one pump is properly installed.
A burette is inserted in the pump and filled with titrant.
The aspiration tube is inserted in the titrant bottle and primed. The dispensing tube
is over the titration beaker.
The standard or sample has been carefully weighed / measured into the titration
beaker.
The electrode(s) and the temperature probe are inserted in the titration beaker.
The desired method is selected and the parameters are set to the optimal values.
6.1.1 Starting a Titration
To start a new titration, press from the main screen.
When a titration begins:
The stirrer will turn on (if detected and enabled).
If the pre-stirring time option is enabled, the sample will be stirred until the set
time elapses (see section
5.5.9 Pre-Titration Stir Time).
If the pre-titration volume option is enabled, the set volume will be dispensed
(see section
5.5.8 Pre-Titration Volume).
The titrator will start the analysis and continue to deliver titrant until the end point is
detected or the titration is terminated.
6.1.2 Suspending a Titration
While a titration is in progress, you can temporarily stop it by pressing . The burette
will stop dispensing titrant.
To continue the titration press .
6.1.3 Viewing the Titration Curve
During a titration, the potentiometric curve and the derivative curve (equivalence point only)
can be displayed on the
Titration Graph
screen, by pressing . The titration report ID
is also displayed inside the graph window.