Property Content Note
Command string MAL
No. of parameters 1
Parameter range 0 … 4
Factory default 4
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
No
Parameter backup With command TDD1
Input Master MAL(P1); No response
Query Master MAL?;
Response WE X crlf 1 character
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
This command defines the access in the parameter menu of the WE:
P1
Access to menu level
0 Information, Scale, OFF
1 Like level 0 and Print
2 Like level 1 and SetPoints, Setup
3 Like level 2 and UART1, UART2, Print Setup, Input Setup, Button setup, Test
4 Like level 3 and Adjust, Factory Adjust