EasyManua.ls Logo

HBM WE2107 - WE2107 Factory Default Curve Setup

HBM WE2107
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
WE2107 - Communication commands
I1820-1.3 en HBM
3.2 Factory default curve
The commands described in this section are used to set up the factory default curve:
D Adjusting the default curve: SZA
, SFA
D Gravitational acceleration correction: ACA, ACU
Setting the characteristic curve
The WE works initially with a factory default curve SZA
, SFA. This factory default calibration
is made with a calibration standard for 0mV/V and 2mV/V (=200000 internal digits). This
factory default curve should not be modified.
A second characteristic curve (LDW
, LWT) is available for the scale adjustment.
The gravitational acceleration correction is then activated via command ACA and ACU, if the
place where the scale was adjusted is not the same as the place of installation and the
gravitational acceleration factors are different.
Setting the factory default curve with SZA
, SFA (absolute value calibration in mV/V)
Action
Command sequence
Enter password, e.g. SPW00000;
Measure input at 0 mV/V SZA(P1);
Measure input at 2mV/V SFA(P2);

Table of Contents

Other manuals for HBM WE2107

Related product manuals