HBM WE2107 - Zero Setting (Initial Zero Setting on Start‐Up)

HBM WE2107
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WE2107 - Communication commands
58
I1820-1.3 enHBM
Zero Setting
(Initial zero setting on start‐up)
Property Content Note
Command string ZSE
No. of parameters 1
Parameter range P1=0 ... 4
Factory default 0 disabled
Reaction time < 15 ms
Password protection Yes
Relevant to legal for
trade
YES
Parameter backup With command TDD1
Input Master ZSE(P1); No response
Query Master ZSE?;
Response WE P1crlf P1 = 1 character
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
When switching on the voltage, during a RESET or after an RES
command, zero setting is
executed in the selected range at standstill after about 2.5 s. Any change to the zero on
start‐up setting range only takes effect after switching on the voltage or after the RES
command.
If there is no standstill or if the gross value is outside the selected limits, zero setting does
not occur. The internal zero memory is always cleared before automatic zeroing. If the gross
value at standstill is within the selected range, the gross value is accepted into the zero
memory. It is not possible to read out the zero memory. The standstill condition will be set
with MDT
. The digit unit (d) relates to the nominal (rated) value (NOV) and the increment
(RES
).
Parameter description:
ZSE0: zeroing disabled
ZSE1: zeroing range "2 % of NOV value
ZSE2: zeroing range "5 % of NOV value
ZSE3: zeroing range "10 % of NOV value
ZSE4: zeroing range "20 % of NOV value
ZSE