43
WE2107 - Communication commands
I1820-1.3 en HBM
Nominal Output Value
(Resolution of the scale characteristic curve)
Property Content Note
Command string NOV
No. of parameters 1
Parameter range P1=100...99999
Factory default 6000
Reaction time <15 ms
Password protection Yes
Relevant to legal for
trade
Yes
Parameter backup With command TDD1
Input Master NOV(P1); No response
Query Master NOV?;
Response WE P1crlf P1 = 6 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The N0V value is used to scale the output value during data output. ASCII data output is
scaled at the factory to 6000. If you require a data output of 2000 digits at max. capacity, for
example, then use this command to set the nominal (rated) value NOV2000;. This scaling
does not change the input parameters.
The tare value is on the NOV curve and is output in this scaling.
Output format measured value at max. capacity
NOV w 100
2‐byte binary NOV value
4‐byte binary NOV value
ASCII NOV value
NOTE
For 2‐byte binary output, the NOV value must be v 30000. Otherwise the measured value
will be output with overflow or underflow (7FFF
H
or 8000
H
; H: hexadecimal). With
NOV30000, the overload reserve is only about 2700 digits.
NOV