Setup of the linearization :
D Use the SPW command to enter your password.
D Set legal for trade switch to LFT=0
D The scale is adjusted (LDW, LWT, NOV, …)
D Switch off the old linearization: LIN1=0, LIN2=0, LIM1=0 LIM2=0,
D The scale is loaded with the first known weight (point1).
D Enter the LIN1 value (weight without comma).
D Query the measurement output (MSV?)
D Enter the LIM1 value (without comma).
D The scale is loaded with the second known weight (point2).
D Enter the LIN2 value (weight without comma).
D Query the measurement output (MSV?)
D Enter the LIM2 value (without comma).
With the input of LIM2 value the WE activates the new linearization curve.
D Store the new values in the EEPROM with TDD1.
Switch OFF the linearization:
Enter the default values:
Enter the password (DPW)
Switch off legal for trade (LFT0; if necessary)
LIN1,0;
LIN2,0;
LIM1,0;
LIM2,0;
TDD1;