WE2107 - Communication commands
I1820-1.3 en HBM
Set Password
(Write enable for all password protected parameters)
Property Content Note
Command string SPW
No. of parameters 1
Parameter range P1= 00000 ... 99999 (5 digits) Must match P1 of
DPW
Factory default 00000
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
Yes
Parameter backup
Input Master SPW(P1); No response
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The command SPW with a correctly entered password (using the command DPW) gives
authorization for data input with all commands. The command SPW with an incorrect
password stops data input for protected commands. A password is not necessary for query.
After an RES or a power‐up, you are again prevented from using the protected commands.
A SPW? query is not permitted.
In the chapter 2.6
there is an overview of the protected commands.