EasyManua.ls Logo

HBM WE2107 - WE2107 External Display Control Commands

HBM WE2107
152 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WE2107 - Communication commands
86
I1820-1.3 enHBM
3.9 Commands for setup the control of an external
display
The commands are used to setup the communication with the external display via the
second serial link.
D Protocol UART2 FUB
D Protocol external display EDP
D Start character EDS
D CRC character EDC
D End character 1 ED1
D End character 2 ED2
Via the second serial link an external display can be connected.
Baud rate: 1200...9600
Parity bit: non / even
Type: RS232
Update rate: 5 / sec.
Protocol: no protocol (just send out)
Hardware – protocol (DTR)
Software – protocol (DC1/DC3/DC4)
The protocol has to be defined in the parameter menu (UART2/FUNCT) or with the
command FUB. The baudrate can only be selected in the parameter menu (there is no
command available)