WE2107 - Communication commands
60
I1820-1.3 enHBM
Measured Signal Value
(Measurement query)
Property Content Note
Command string MSV?
No. of parameters
Parameter range
Factory default
Reaction time < 15 ms
Password protection No
Relevant to legal for
trade
No
Parameter backup No data to back up
Query Master MSV?;
Response WE See description
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
As previously defined, the measured value is output in ASCII or binary format (see
commands COF
, NOV and RSN). The maximum extent of the measured values is:
For 2‐byte data output: integer ±32767
For 4‐byte data output: long integer ±399999
For ASCII data output: ASCII ±399999
The length of output depends on the output format (see COF
command).
The output format for a measured value must be defined before the measurement run. The
measured value is output in relation to the particular measuring range (NOV
). The measured
value can be a net or a gross measured value (TAS
).
MSV