WE2107 - Communication commands
I1820-1.3 en HBM
Sensor Zero Adjust
(factory default curve zero point)
Property Content Note
Command string SZA
No. of parameters 1
Parameter range P1=0...+
399999
Factory default Adjustment to 0 mV/V
Reaction time <15 ms on input or query
Password protection Yes
Relevant to legal for
trade
Yes
Parameter backup With command TDD1; after input of SFA
Input Master SZA(P1); No response
Query Master SZA?;
Response WE P1crlf
(P1= 6‐digit plus sign)
P1=
7 characters
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
For an input signal of 0mV/V, the output value 0 digits is assigned to the internal measured
value.
Parameter description:
For a query, the value is output ±6‐digit (e.g. ‐000246 CRLF).
The curve is disabled at SZA=0 and SFA=200000.
To perform the electronic adjustment see command SFA
.