WE2107 - Communication commands
74
I1820-1.3 enHBM
Restart
(Reset electronic)
Property Content Note
Command string RES
No. of parameters ‐
Parameter range ‐
Factory default ‐
Reaction time < 4 s
Password protection No
Relevant to legal for
trade
No
Parameter backup ‐
Input Master RES; No response
Note: ( ) required parameters, < > optional parameters for parameter input
Function:
The RES command initiates device start‐up (warm start). This command does not generate a
response. All the parameters are set in the same way as they were saved with the last TDD1
command, that is to say, the EEPROM values are transferred to the RAM.
A RES? query is not permitted.
RES